型号:Genius3000




检测仪实验数据

               
含量单位: PPM EC681K
序号次数 工作曲线  As Br Cd Cr Hg Pb
1 PE 30.5568 775.628 140.4 24.194 103.15 100.973
2 PE 30.4782 770.526 140.08 24.764 100.52 99.2205
3 PE 31.5777 766.347 135.44 24.294 98.674 102.973
4 PE 29.9576 769.997 136.38 25.216 100.2 100.845
5 PE 30.3706 769.872 133.74 24.216 101.08 102.973
6 PE 30.6149 771.922 134.2 23.975 100.79 103.976
7 PE 30.391 771.009 135.32 24.5 100.99 101.487
8 PE 30.3195 768.284 136.61 25.167 99.502 99.5987
9 PE 30.6208 769.656 135.56 24.196 103.65 97.705
10 PE 30.8942 768.275 133.53 24.699 98.026 103.761
含量真实值 29.1 770 137 23.7 98 100
含量平均值 30.5781 770.152 136.13 24.522 100.66 101.351
测量标准偏差Sn 0.40532 2.363 2.2759 0.4059 1.6701 1.9819
3倍Sn值 1.21595 7.08901 6.8276 1.2178 5.0104 5.94569
测量差Rppm 1.62  9.28  6.87  1.24  5.62  6.27 
测量相对标准偏差 1.33% 0.31% 1.67% 1.66% 1.66% 1.96%
示值误差ε(%) 5.08% 0.02% -0.64% 3.47% 2.71% 1.35%

检测
1.配备新SDD探测器,探测器分辩率zui低<139eV,对于微量有害元素可以进行更加的测试。
2.配备高清晰摄像头,可以更加准确的进行点定位测试。对于针脚之类的微小样品进行测试。
3.全新软件,算法更强,功能更多。提高了有害元素的检测精度。
4.常压充氦气系统,大大提高了卤素(尤其是Cl元素)的测试精度。

应用领域
1、可用于体积构件的ROHS及有害元素测试
2、电子元件与部件的材料现场检测
3、各种玩具、文具、儿童用品和礼品的现场抽检。
4、包装材料进行测试与验证。
5、各类电池进行有害物质测试与验证。
6、电镀液中有害元素的测试
7、布料、鞋材等材料中有害元素的检测。
8、饰、饰品中的有害元素测试。