Enhancements to 5700A/5720A ac voltage output capability provided by the 572:
- Frequency limits at higher voltage increase to 100 kHz at 750 V, 30 kHz at 1100 V
- Load limits are to 70 mA for frequencies above 5 kHz, and to 50 mA for frequencies less than 5 kHz
- Capacitive load limits are increased to 1000 pF
Voltage output from the 572 is available at the 5700A Series Calibrators' front or rear binding posts. This eliminates the need to move cables during a procedure that requires amplified as well as standard calibrator outputs.福禄克572 Amplifier
The
572 front or rear panel OUTPUT binding posts are o
Model 572 operating functions and ranges:
- AC voltage: 220 V to 1100 V rms up to 70 mA (50 mA < 5 kHz), 40 Hz to 30 kHz; 220 to 750V rms up to 70 mA, 30 kHz to 100 kHz
- DC current: 0 A to ±11 A
- AC current: 1 A to 11 A rms, 40 Hz to 10 kHz